The hard part of digital-health regulation in the EU is not any single regime. It is that four of them now apply to the same piece of software at once, and each was drafted as if the others did not exist.
A firm shipping a diagnostic or clinical-decision app in the EU used to have one question to answer: is this a medical device? That question is still first, but it is no longer last. Once the software crosses the device line, the Medical Device Regulation, the AI Act, EUDAMED registration, the European Health Data Space and a cybersecurity regime all attach to it. They stack; they do not substitute. This is a survey of where each one bites, written for the CTO or founder who has to decide, before the next release, which obligations their software has already triggered.
The moment software becomes a device
The line is drawn by the Medical Device Regulation (Regulation (EU) 2017/745, the MDR) and its in-vitro sibling (Regulation (EU) 2017/746, the IVDR). Software has a medical purpose the moment it is intended to diagnose, prevent, monitor, predict, treat or alleviate disease. General wellness, fitness and lifestyle software sits outside; the instant a claim tips into clinical decision-making, it is in.
Free · 4 minutes
Do you know where AI is already being used in your business — and what it can see?
Fourteen questions on shadow AI, data exposure, oversight, and governance debt — the gap between how fast AI is arriving and how much control you have over it. Banded finding on screen, full sheet by email.
What surprises most teams is the class the software lands in. MDR Annex VIII, Rule 11 governs software, and it is deliberately upward-biased. Software intended to provide information used to take decisions with a diagnostic or therapeutic purpose is Class IIa by default. If a wrong decision could cause a serious deterioration in health or lead to a surgical intervention, it is Class IIb. If it could cause death or an irreversible deterioration, it is Class III. Only software that does none of this falls to Class I. The practical effect is that most genuinely useful clinical software is at least Class IIa, and Class IIa is the threshold at which a notified body must be involved. The IVDR classifies its software differently, on an A-to-D scale keyed to the risk of a wrong result, but the direction of travel is the same. The detail of that jump, and why so many firms misjudge it, is worth reading in full in our note on SaMD classification under MDR/IVDR Rule 11.
The AI Act stacks on top, it does not replace
If the software learns, or uses a model, the AI Act (Regulation (EU) 2024/1689) attaches as a second, parallel regime. Its Article 6(1) is the hinge: an AI system is high-risk where it is, or is a safety component of, a product covered by the Union harmonisation legislation in the Act’s Annex I, and that product must undergo third-party conformity assessment. The MDR is Annex I legislation. Any medical-device software at Class IIa or above already needs a notified body. So a learning diagnostic system is, almost automatically, both a regulated device and a high-risk AI system, with no separate risk assessment required to get there.
The regulators saw the duplication coming. MDCG 2025-6 sets out the intended interplay: one combined conformity assessment, carried out where possible by the same notified body, against a single integrated technical file, with the AI Act requirements folded into the existing MDR procedure rather than run in parallel. That is the theory. In practice it means the AI Act’s obligations — data governance, logging, human oversight, robustness and technical documentation — become extra chapters in a device file you already have to maintain. Model provenance is one of them, and it is the chapter most teams have not started; our piece on the AI-BOM covers what that documentation has to contain.
The timing is genuinely in flux. Under the AI Act as adopted, obligations for high-risk systems that are Annex I products apply from 2 August 2027. The Commission’s Digital Omnibus reached political agreement in May 2026 to defer that to 2 August 2028, with Annex III stand-alone systems moving to 2 December 2027. Those new dates are not yet in the Official Journal, so treat them as provisional and verify against the final text before you plan around them.
EUDAMED: the registration you can no longer defer
EUDAMED, the European database on medical devices, spent years as a voluntary system nobody had to use. That ended with Commission Decision (EU) 2025/2371, published in November 2025. Four of its modules — actor registration, UDI/device registration, notified bodies and certificates, and market surveillance — become mandatory from 28 May 2026. Devices already on the market must be registered by 28 November 2026, and notified-body certificates by 28 May 2027. The vigilance and clinical-investigation modules remain under development, so those obligations follow later. If your product is a regulated device, you now need a Basic UDI-DI, actor registration and a device entry in a live public database, on a fixed clock. This is administrative, not clever, and it is exactly the sort of deadline that slips because no engineer owns it.
EHDS: the data-format obligation nobody costed
The European Health Data Space (Regulation (EU) 2025/327) entered into force on 26 March 2025 and applies in phases through the late 2020s — primary-use provisions from March 2027, EHR-system conformity requirements from March 2028, and the secondary-use framework from March 2029. For a device maker the load-bearing point is technical, not legal. EHDS builds on a common European exchange format anchored in HL7 FHIR, with the International Patient Summary profile and terminologies such as SNOMED CT and LOINC. Software that produces or consumes electronic health records will be expected to speak that format, and EHR systems face their own conformity assessment. Retrofitting a proprietary data model onto FHIR is a project, not a config flag, and it is one most digital-health roadmaps have not costed.
Cybersecurity: MDR first, the CRA at the edges
The instinct is to assume the Cyber Resilience Act (Regulation (EU) 2024/2847) now governs device security. Mostly it does not. The CRA carves out products already covered by the MDR and IVDR to avoid double regulation, so for a regulated device the security obligations live in the MDR’s own general safety and performance requirements, interpreted through MDCG 2019-16. The CRA still catches the digital-health products that sit outside the device perimeter — wellness apps, general-purpose components, connected accessories — and its main obligations apply from 11 December 2027. Knowing which regime owns a given component, and running one SBOM-backed programme across both, is the practical task; we set it out in our note on meeting MDCG 2019-16 and the CRA together.
What this actually means for the build
Read together, the regimes share evidence more than they share text. The same technical file supports MDR classification, AI Act conformity and, increasingly, EHDS conformity; the same UDI threads through EUDAMED; the same SBOM underpins both MDR and CRA cybersecurity. A firm that treats these as five separate compliance projects will build five overlapping evidence sets and reconcile none of them. A firm that maps them onto one device dossier does the work once. The regulators have started designing for the second approach. Most digital-health teams are still resourced for the first.
The question is no longer whether your software is a medical device. It is how many regimes noticed at the same moment it became one.
Most technology problems are not technology problems. They are control problems.
The systems exist. The investment has been made. The question is whether leadership can understand, direct, evidence, and sustain what those systems produce. Find out where control exists — and where it only appears to.
Full Governance by Sixteen Pillars
Govern your business. Prove your compliance.
A board assurance cockpit for EU-regulated financial firms — tamper-evident, hash-chained proof of governance across DORA, GDPR, NIS2, ISO 27001, the EU AI Act and MiCA. In development.
See what's coming